null
AirLife

Salter PRO2 Check Elite Oxygen Analyzer – Ultrasonic Sensor, No Fuel Cell to Replace

(No reviews yet) Write a Review
$1,334.99
SKU:
PRO2ELITE-1
MPN:
PRO2ELITE-1
Shipping:
FREE standard shipping (contiguous US)
  • Salter PRO2 Check Elite Oxygen Analyzer – Ultrasonic Sensor, No Fuel Cell to Replace
  • Salter PRO2 Check Elite Oxygen Analyzer – Ultrasonic Sensor, No Fuel Cell to Replace
$1,334.99
Fast Delivery
Fast Delivery
Most ship in 1–2 days
Shipping
$50+ Ship Free
Flat $6.95
30-Day Returns
30-Day Returns
No questions asked
Secure Payment
Secure Payment
100% safe checkout

Description

The Salter PRO2 Check Elite Oxygen Analyzer gives home care providers one handheld instrument for the three field checks every oxygen concentrator service visit requires. It measures oxygen concentration, verifies flow, and reads outlet pressure at the machine, and it shows each reading on its LCD in about 10 seconds.

  • – Ultrasonic oxygen sensing with no fuel cell to replace
  • – Three modes in one unit: concentration, flow, and outlet pressure
  • – Factory calibrated, with a pure oxygen verification check built in
  • – Runs on one 9-volt battery and shuts off after about 5 idle minutes

Salter PRO2 Check Elite Oxygen Analyzer – Ultrasonic Sensor, No Fuel Cell to Replace

A concentrator can sound healthy while its sieve beds fade and its delivered oxygen drops well below the level on the order. The PRO2 Check Elite was built to catch that drift in the home, using the same ultrasonic sensing technology that concentrators themselves use to trigger their low oxygen alarms. Salter Labs combined that sensor with built-in flow and pressure measurement, so one instrument checks the three output parameters that decide whether a concentrator is doing its job. In its default concentrator mode, the unit compensates automatically for the argon that passes through molecular sieve beds, which keeps readings true to what the patient actually receives. It can also spot check liquid oxygen vessels and gaseous oxygen cylinders. Two anodized aluminum inlet ports on top of the housing connect with the supplied 2 ft tubing, and the whole kit rides in a carrying case with a strap.

Features and Benefits

Measurement Performance

  • Oxygen sensing: ultrasonic sensor measures concentration from 20.9% to 100% with no fuel cell to replace
  • Concentration accuracy: readings accurate to ±2%, with a response time of about 10 seconds
  • Flow measurement: 0 to 10 LPM range with 0.1 LPM resolution for verifying the concentrator's own flow control
  • Pressure measurement: 0 to 25 PSIG range, displayed in PSI or kPa

Field Use

  • Argon compensation: concentrator mode corrects automatically for argon passing through the sieve beds
  • Battery power: one standard 9-volt alkaline battery, factory installed, with a low battery indicator on the LCD
  • Automatic shut-off: the unit powers down after about 5 minutes without use to preserve battery life
  • Calibration checks: pure oxygen verification plus one-step and two-step calibration procedures are built into the controls

Product Specifications

Measurement Ranges

  • Concentration range: 20.9% to 100% oxygen
  • Concentration accuracy: ±2%, with linearity of ±2% of full scale
  • Flow range: 0 to 10 LPM, accuracy ±0.3 LPM (±3% of full scale)
  • Pressure range: 0 to 25 PSIG (0 to 172 kPa), accuracy ±1% of full scale

Product Details

  • Dimensions: 3.6 x 5.75 x 1.29 in (9.14 x 14.6 x 3.28 cm)
  • Weight: 9 oz (255 g)
  • Connectors: two 1/4 in anodized aluminum hose barbs
  • Case material: ABS plastic
  • Power source: one 9-volt alkaline battery, factory installed
  • Operating temperature: 32 to 105 °F (0 to 41 °C)
  • Package contents: indicator unit, 9-volt battery, 2 ft connecting tubing, carrying case and strap, operation manual
  • Packaging: CS / 1
  • Warranty: Salter Labs 2-year warranty to the original purchaser, covering repair or replacement if the unit becomes inoperable within two years of purchase

Browse the professional use range for more clinical testing equipment, or see the full AirLife range for this manufacturer's other respiratory products.

Why Choose the Salter PRO2 Check Elite?

A concentrator that is quietly delivering 82% oxygen instead of 93% looks fine from the outside, and the patient feels the difference long before anyone measures it. This analyzer is how a provider proves output on the spot, in one visit, with one instrument.

  • – One instrument replaces a separate analyzer, flow meter, and pressure gauge in the service bag
  • – Ultrasonic sensing responds in about 10 seconds and never needs a replacement fuel cell
  • – Calibration holds through battery changes, so the unit is ready when a visit runs long
  • – At 9 oz with its own case and strap, it is built for route work, not the bench

Manuals and Troubleshooting Guides

Salter PRO2 Check Elite Oxygen Analyzer – Operation Manual

Frequently Asked Questions

What does the PRO2 Check Elite measure?

It measures the three outputs that decide whether an oxygen concentrator is working correctly: oxygen concentration, flow rate, and outlet pressure. It can also spot check liquid oxygen vessels and gaseous oxygen cylinders.

Is a prescription needed to buy this analyzer?

No. The PRO2 Check Elite is a service instrument, and Salter Labs states in its operation manual that the instrument has no clinical diagnostic or therapeutic functions. Providers use it to check equipment output, not to treat or diagnose patients.

How accurate are its readings?

Oxygen concentration readings are accurate to ±2% across the 20.9% to 100% range. Flow readings are accurate to ±0.3 LPM across 0 to 10 LPM, and pressure readings are accurate to ±1% of full scale across 0 to 25 PSIG.

Does the unit need routine calibration?

It leaves the factory calibrated, and calibration holds even when the battery is changed. Accuracy can be verified at any time against a USP oxygen source above 99.0%, and the manual documents one-step and two-step calibration procedures for organizations whose policies require periodic calibration.

What battery does it use, and how long does it last?

It runs on one standard 9-volt alkaline battery, which is factory installed. The unit shuts itself off after about 5 minutes without use to preserve battery life, and the LCD shows a LO BATT warning when the battery runs low.

What warranty covers the analyzer?

Salter Labs covers the unit for two years from the date of purchase. If it becomes inoperable in that period for any reason other than abusive use, a dead or defective battery, or water intrusion, Salter Labs will repair or replace it at its option.

View AllClose

Warranty Information

Salter Labs 2-year warranty to the original purchaser: if the unit becomes inoperable within two years of the date of purchase for any reason other than abusive use or handling, a dead or defective battery, or water intrusion, Salter Labs will, at its option, repair or replace the unit at no charge for parts or labor.
View AllClose

Additional Information

Packaging:
CS / 1
View AllClose

0 Reviews

View AllClose

Explore our wide range of medical products, enjoy hassle-free shopping, and take advantage of our exclusive offers. Free shipping on orders over $50! Need assistance? Contact our friendly customer support team at (800) 494-5750. Happy shopping!